A handheld veterinary multigas monitor, the veterinary version of the AneSure-H. It measures CO2, nitrous oxide (N2O) and anesthetic agents by sidestream sampling, with optional SpO2 and pulse rate.
Sampling flow 50–250 ml/min, selectable.
Measures nitrous oxide and halothane, enflurane, isoflurane, sevoflurane or desflurane.
Inserting the filter starts the sampling pump automatically; removing it stops the pump.
Automatically compensates for changes in atmospheric pressure.
SpO2 and pulse rate measurement available as an option.
| Technique | NDIR |
| Power | AC 100–240 V, 50/60 Hz; DC 3.7 V; lithium battery 3.6 V, 3800 mAh |
| Battery working time / charging | Up to 8 hours; charging time 4 hours |
| Operating environment | 5–50 °C; humidity <93%; 70–120 kPa |
| Size / weight | 72 × 155 × 38 mm (W × H × D); 360 g |
| Display | 3.5 inch TFT |
| CO2 range | 0–19.7%, 0–150 mmHg or 0–20 kPa |
| CO2 accuracy | 0–40 mmHg: ±2 mmHg; 41–70 mmHg: ±5% of reading; 71–100 mmHg: ±8% of reading; 101–150 mmHg: ±10% of reading |
| Respiratory rate | 3–150 breaths/min; 1% of reading or ±1 breath/min |
| N2O | 0–100%; ±(2 vol% + 2% of reading) |
| Halothane, enflurane, isoflurane | 0–12%; 0–8%: ±(0.15 vol% + 5% of reading); 8–12%: unspecified |
| Sevoflurane | 0–15%; 0–10%: ±(0.15 vol% + 5% of reading); 10–15%: unspecified |
| Desflurane | 0–25%; 0–18%: ±(0.15 vol% + 5% of reading); 18–25%: unspecified |
| Sampling flow | 50–250 ml/min, selectable |
| SpO2 (optional) | 0–100%; ±3% at 50–69%, ±2% at 70–100% |
| Pulse rate (optional) | 30–250 bpm; ±2 bpm or ±2% of reading, whichever is greater |
| EtCO2 update / trends | Every breath; 24-hour trends |
| Product | Item | Description | Use / replacement |
|---|---|---|---|
![]() | Filter T4F | Water filter for the sidestream inlet; 0.2 µm membrane; female Luer | Single patient. Replace every 2 months, and immediately if sticky or viscous material gets into the filter |
![]() | Sampling tube | 2.4 m; adult or pediatric; connects the filter or water trap to the T-adapter | Disposable |
![]() | T-adapter | Airway T-piece (662531) for intubated patients; tighten the connections so they do not leak | Reusable after wiping with 75% alcohol |
![]() | SpO2 probe (optional) | For use with compatible SpO2 measuring devices | Measure preferably on the index finger; the middle or ring finger can also be used |