A built-in (OEM) sidestream end-tidal CO2 module for intubated and non-intubated adult, pediatric and neonatal patients. It works with the reusable T2 water trap.
Temperature and pressure compensation, automatic timed and drift self-calibration and full-range factory calibration ensure accurate measurement over 0–20% CO2; automatic barometric compensation over 400–860 mmHg.
Low power consumption, suitable for battery-powered use.
Redundant design of the gas path, hardware and software keeps the module stable over the long term. The low dead-space T2 water trap is reusable, lowering running costs.
RS-232 or TTL (3.3 V or 5 V), chosen when ordering; custom protocols on request.
Detects whether the water trap is inserted and indicates occlusion, leaks or pump faults by an LED on the water trap holder.
| Sensor / technique | Dual-channel infrared sensor; non-dispersive infrared (NDIR) |
| Operating conditions | 5–50 °C, 10–90% RH non-condensing; 55–115 kPa |
| Storage | -40–70 °C, <90% RH non-condensing |
| Power | 5 V ±5% (max ripple 200 mVp-p); typically 120 mA, 280 mA during offset calibration |
| Interface | RS-232 or TTL (3.3 V/5 V), Standard E-8-1, 38.4 kBaud, 10 ms data interval, or customised |
| CO2 range / resolution | 0–19.7% (0–150 mmHg or 0–20 kPa); 0.1 mmHg |
| CO2 accuracy | 0–40 mmHg: ±2 mmHg; 41–70 mmHg: ±5% of reading; 71–100 mmHg: ±8% of reading; 101–150 mmHg: ±10% of reading |
| Respiratory rate | 2–150 breaths/min; 1% ±1 breath/min |
| Gas compensation | N2O, O2, agent (host supplied) |
| Barometric compensation | 400–860 mmHg, automatic |
| Warm-up | 97% of designed accuracy within 45 s; full accuracy within 2 min |
| Rise time (t10–90%) | 200 ms |
| Delay time | <3 s at 100 ml/min with adult water trap and 2.4 m sampling line |
| Sampling flow | Default 120 ml/min ±10%; adjustable 50–200 ml/min by host command |
| Water trap emptying | Every 24 h at 100 ml/min, every 16 h at 150 ml/min (5–50 °C, 10–90% RH) |
| Product | Item | Description | Use / replacement |
|---|---|---|---|
![]() | Water trap T2 | Reusable water trap with 10 ml bottle; 0.2 µm membrane | Empty the bottle when it is two-thirds full; clean only the water bottle, with 75% alcohol |
![]() | Sampling tube | 2.4 m; adult or pediatric; connects the filter or water trap to the T-adapter | Disposable |
![]() | Nasal sampling cannula | Adult/pediatric or infant | Disposable |
![]() | CO2/O2 nasal cannula | Samples CO2 and delivers oxygen; adult/pediatric | Disposable |
![]() | T-adapter | Airway T-piece (662531) for intubated patients; tighten the connections so they do not leak | Reusable after wiping with 75% alcohol |