A portable, handheld capnograph dedicated to veterinary use. It monitors EtCO2 (real-time values and waveform) and respiratory rate, with optional SpO2 and pulse rate.
Monitors EtCO2/CO2 and respiratory rate; SpO2 and pulse rate optional.
Kingst's EtCO2 technology of infrared gas analysis.
Temperature and pressure compensation and a zero readjustment system ensure accuracy and stability; balance gas compensation minimises the influence of other gases.
Sidestream measurement, for intubated and non-intubated patients. Sampling flow 50–250 cc/min, selectable.
Audible and visual alarms for limits and apnea; 24-hour trends kept after power-off. Units kPa, mmHg or % selectable.
Connects to a mobile device by Bluetooth for real-time data transmission; stored data can be transferred to a PC by USB for saving or printing reports. The CapnoCloud app allows remote monitoring of pets in the operating room.
| Power | AC 100–250 V, 50/60 Hz; battery DC 3.6 V / 3800 mAh; AC power consumption ≤5 VA |
| Battery working time | Up to 8 hours after full charge |
| Operating environment | 5–40 °C; 30–75% RH, non-condensing; 55–115 kPa |
| Size | 70 × 160 × 44 mm (W × H × D) |
| Net weight | 400 g (lithium battery included) |
| Display | 2.8 inch, colour TFT |
| CO2 unit | mmHg, kPa, % (selectable) |
| CO2 range | 0–150 mmHg |
| CO2 accuracy | 0–40 mmHg: ±2 mmHg; 41–70 mmHg: ±5% of reading; 71–100 mmHg: ±8% of reading; 101–150 mmHg: ±10% of reading |
| Respiratory rate | 3–150 breaths/min; ±1% or ±1 breath/min, whichever is greater |
| Measurement | Sidestream measurement, for intubated and non-intubated patients |
| Sampling flow | 50–250 cc/min, selectable |
| SpO2 (optional) | 0–100%; ±3% at 50–69%, ±2% at 70–100% |
| Pulse rate (optional) | 30–250 bpm; ±2 bpm or ±2% of reading, whichever is greater |
| Trends | 24 hours |
| Alarms | Visual and audible: check adapter, no adapter, apnea, low battery, EtCO2/RR above or below limits |
| Product | Item | Description | Use / replacement |
|---|---|---|---|
![]() | Filter T4F | Water filter for the sidestream inlet; 0.2 µm membrane; female Luer | Single patient. Replace every 2 months, and immediately if sticky or viscous material gets into the filter |
![]() | Sampling tube | 2.4 m; adult or pediatric; connects the filter or water trap to the T-adapter | Disposable |
![]() | Nasal sampling cannula | Adult/pediatric or infant | Disposable |
![]() | T-adapter | Airway T-piece (662531) for intubated patients; tighten the connections so they do not leak | Reusable after wiping with 75% alcohol |
![]() | SpO2 probe (optional) | For use with compatible SpO2 measuring devices | Measure preferably on the index finger; the middle or ring finger can also be used |