A portable, palm-size veterinary multigas monitor, the veterinary version of the AneSure mini. It monitors EtCO2 (real-time value and waveform), FiCO2, nitrous oxide, one of five anesthetic agents (halothane, enflurane, isoflurane, sevoflurane or desflurane, selected manually), MAC and respiratory rate.
CO2, N2O and the selected anesthetic agent with MAC, in an 80 g unit.
Measures at the endotracheal tube through the mA3/mA3N airway adapter.
Turn on the unit and it is ready to use, with no need to calibrate.
First EtCO2 value in 5 seconds, accurate readings within 15 seconds.
Select the anesthetic agent to be monitored manually.
| Power | AC 220 V 50 Hz; battery DC 3.7 V / 1.2 Ah; AC power consumption ≤10 VA |
| Battery working time | 6–8 hours after full charge |
| Operating environment | -5–40 °C; 10–90% RH, non-condensing; 55–115 kPa |
| Size | 38 × 42 × 44 mm (W × H × D) |
| Net weight | 80 g (lithium battery and adult adapter included) |
| Display | 1.5 inch, colour TFT |
| CO2 unit | mmHg, kPa, % (selectable) |
| CO2 range | 0–150 mmHg |
| CO2 accuracy | 0–40 mmHg: ±2 mmHg; 41–70 mmHg: ±5% of reading; 71–100 mmHg: ±8% of reading; 101–150 mmHg: ±10% of reading |
| Respiratory rate | 2–150 breaths/min; ±1% or ±1 breath/min, whichever is greater |
| N2O | 0–100%; ±(2 vol% + 2% of reading) |
| Halothane, enflurane, isoflurane | 0–12%; 0–8%: ±(0.15 vol% + 5% of reading); 8–12%: unspecified |
| Sevoflurane | 0–15%; 0–10%: ±(0.15 vol% + 5% of reading); 10–15%: unspecified |
| Desflurane | 0–25%; 0–18%: ±(0.15 vol% + 5% of reading); 18–25%: unspecified |
| EtCO2 update / trends | Every breath; 24-hour trends |
| Adapter | Patient type | Dead space | Use |
|---|---|---|---|
| mA3 | Adult / pediatric | ≤5 ml | Single-use |
| mA3N | Infant | ≤1 ml | Single-use |