English

京ICP备18007245号-1 <br/>药品医疗器械网络信息服务备案:(京)网药械信息备字(2022)第00481号<br/>Beijing Kingst Commercial & Trade Co .,Ltd.
京ICP备18007245号-1 <br/>药品医疗器械网络信息服务备案:(京)网药械信息备字(2022)第00481号<br/>Beijing Kingst Commercial & Trade Co .,Ltd.

Two Different ETCO₂ Monitoring Methods: Mainstream and Sidestream

Two Different ETCO₂ Monitoring Methods: Mainstream and Sidestream

ETCO₂ (End-Tidal Carbon Dioxide) monitoring is an essential tool that provides real-time information about pulmonary ventilation efficiency. By measuring the carbon dioxide concentration in exhaled gas, it offers clinicians immediate insight into a patient’s respiratory status. Compared with traditional pulse oximetry, ETCO₂ monitoring can detect respiratory abnormalities earlier. It is widely used in anesthesia, surgery, ICU, and emergency settings.


1. Concept and Physiological Significance of ETCO₂

  • Definition: ETCO₂ is the highest CO₂ concentration in exhaled gas at the end of expiration.

  • Physiological Significance:

    • Reflects alveolar ventilation: ETCO₂ directly indicates alveolar ventilation efficiency. Low ETCO₂ suggests hypoventilation, while rapid ventilation can also lower ETCO₂ values.

    • Indirectly reflects blood CO₂ levels: ETCO₂ is usually slightly lower than arterial CO₂ (PaCO₂), and continuous monitoring provides trends in blood CO₂ levels.

    • Early detection of respiratory abnormalities: ETCO₂ can indicate hypoventilation, hyperventilation, apnea, or airway obstruction, serving as an early warning, especially in anesthesia and critical care.

    • Clinical reference: ETCO₂ waveforms show the completeness of the respiratory cycle, and the changes at the end of exhalation and inhalation help clinicians evaluate airway patency and breathing patterns.


2. Two ETCO₂ Measurement Methods

ETCO₂ monitoring primarily uses Mainstream and Sidestream methods. The principles, installation, and operational characteristics differ between the two.


2.1 Mainstream Measurement

  • Principle: A sensor is directly installed in the breathing circuit. As gas flows through, the optical or infrared sensor measures CO₂ concentration directly, providing real-time waveforms and values.

  • Advantages:

    • High real-time response, no waveform delay

    • High measurement accuracy

    • Less affected by environmental factors such as humidity and temperature

  • Disadvantages:

    • Adds slight weight to the breathing circuit

    • Requires secure connection to the circuit to ensure accuracy

  • Operational Notes:

    • Calibration (zeroing) is required when replacing the breathing circuit adapter

Diagram:

图片4.png

2.2 Sidestream Measurement

  • Principle: A small internal pump in the monitor draws a sample of gas from the patient’s airway through a thin tubing into the instrument, where CO₂ concentration is measured and waveforms and values are displayed.

  • Advantages:

    • Adjustable sampling flow ensures stable gas measurement

    • Lightweight and easy to install, with minimal added resistance to the breathing circuit

    • Does not add weight directly to the breathing circuit, suitable for mobile or space-limited applications

  • Disadvantages:

    • Slight waveform delay may occur

    • Measurement can be affected by circuit dead space; insufficient sample gas can lead to lower values

  • Usage Requirements:

    • Requires tubing connection between the patient’s airway and the monitor

  • Operational Notes:

    • Regularly replace the filter to prevent blockage that could affect measurement accuracy

Diagram:

图片5.png


Two Different ETCO₂ Monitoring Methods: Mainstream and Sidestream